Core workflow and compatibility
The TrueAmp Library Preparation Kit supports a broad range of DNA input amounts, from as low as 100 pg up to 1 µg. This flexibility enables reliable library preparation across low-input and high-input sample types.
No, the TrueAmp Library Preparation Kit is not compatible with bisulfite conversion or EM-seq workflows. The enzymatic steps in the TrueAmp Library Preparation workflow include DNA repair, which can alter or remove native methylation signatures.
For methylation-sensitive applications such as bisulfite or EM-seq, we recommend using mechanical shearing (e.g., sonication) followed by the Twist cfDNA Library Preparation Kit, which is better suited for preserving methylation information.
The TrueAmp Library Preparation Kit is not recommended for cfDNA workflows. Since cfDNA is already fragmented, the enzymatic fragmentation step is unnecessary. While it is possible to skip fragmentation and proceed with end repair and dA-tailing, the one-pot workflow may not deliver optimal conversion efficiency for these short fragments.
For best performance and highest library conversion from cfDNA, we recommend using the Twist cfDNA Library Preparation Kit, which is specifically optimized for cell-free DNA.
Yes, the TrueAmp Library Preparation Kit performs well with degraded and FFPE-derived DNA. The workflow includes DNA repair activity, which can help restore damaged bases commonly found in FFPE samples.
As a result, users often observe improved library yield and recovery compared to workflows without repair. Some optimization (e.g., fragmentation time, PCR cycles) may be required depending on sample quality.
Yes. The TrueAmp fragmentation chemistry is broadly compatible with a range of double-stranded DNA inputs, including genomic DNA (gDNA), plasmids, and PCR-derived amplicons. Under consistent reaction conditions, high-quality DNA substrates, regardless of origin, typically yield comparable fragment size distributions, indicating minimal sequence or topology-dependent bias.
During development, fragmentation performance was evaluated using DNA ladders spanning a wide range of fragment sizes. This study demonstrates that the enzymatic fragmentation preferentially processes larger DNA molecules while maintaining representation of shorter fragments, resulting in a controlled size distribution without significant loss of diversity across the input population.
Yes, the TrueAmp Library Preparation Kit is well suited for targeted enrichment workflows, including hybrid capture applications. It generates high library yields and promotes uniform molecule conversion, which supports efficient and consistent downstream target capture performance.
Yes, the TrueAmp Library Preparation Kit is compatible with liquid handling automation platforms such as Hamilton systems. The streamlined workflow, reduced number of steps, and minimal hands-on time make it well suited for automation. Minor optimization (e.g., mixing, bead handling, or liquid classes) may be required depending on the platform and configuration.
Reaction setup
Fragment size can be controlled by adjusting fragmentation reaction temperature and incubation time. Increasing either parameter generally results in shorter DNA fragments, while reducing them produces longer fragments.
Because fragmentation performance can vary depending on thermocycler calibration, DNA input quality, and storage buffer composition, we recommend performing a small design of experiments (DOE) with your specific samples to determine optimal conditions prior to scaling up.
For additional guidance and example data, please refer to the Twist TrueAmp Library Preparation Kit technical note on insert size optimization.
Input DNA quality directly affects fragment size. High-quality, intact DNA produces predictable insert sizes, while degraded or pre-sheared DNA (e.g., FFPE) typically results in shorter fragments under the same conditions. In these cases, reducing fragmentation time or temperature and applying size selection may help achieve the desired insert size.
Yes, fragmented and end-prepped DNA can be safely stored at either 4 °C or -20 °C (e.g., overnight) prior to ligation without any appreciable impact on library preparation performance as compared to samples processed immediately. This has been validated across technical replicates (n=3), where samples were stored under these conditions before ligation and showed consistent results.
Yes. Vortexing the 10X Twist Fragmentation Enzyme Mix prior to use is important to ensure the solution is fully homogeneous. Failure to properly mix the enzyme solution may lead to variability in fragmentation performance and inconsistent results.
The TrueAmp Library Preparation Kit is designed for flexibility and is compatible with a wide range of 3′ T-overhang adapters. Note that adapter quality impacts overall library preparation efficiency. For optimal performance, we recommend using validated Twist adapter systems:
- Twist Universal Adapter System (SKU: 101308)
- Requires PCR amplification with indexed primers to complete library construction.
- Twist UMI Adapter System (SKU: 105041)
- Requires PCR amplification with indexed primers.
- Enables incorporation of Unique Molecular Identifiers (UMIs) for improved error correction and duplicate handling.
- Twist Full-Length UDI Adapters (SKU: 107376)
- Do not require PCR for index incorporation, as full-length P5/P7 sequences and indices are already present.
- Optional PCR may still be performed for library amplification if additional yield is needed.
All three systems are fully compatible with the TrueAmp workflow. Selection should be based on your application needs (e.g., UMI requirements, workflow, or PCR-free indexing preference).
Adapter input should be adjusted based on DNA input amount and fragment size to maintain sufficient adapter-to-insert molar excess for efficient ligation while minimizing adapter-dimer formation. Twist Universal Adapters are supplied at 10 µM and recommended conditions in the protocol are designed to maintain an adapter-to-insert molar ratio of approximately ≥150:1 for robust ligation efficiency across a broad range of inputs. Too little adapter may reduce library conversion and yield, while excessive adapter can increase adapter-dimer carryover, particularly in low-input samples.
If optimization is required, titrate adapter input while monitoring final library yield and adapter-dimer levels. The optimal condition is the lowest adapter amount that maintains strong library yield with minimal adapter dimer.
For ultra-low input samples (<1 ng), additional optimization may be required to balance ligation efficiency against adapter-dimer formation, as reduced insert mass increases the likelihood of adapter self-ligation products. Conversely, for higher DNA input amounts, increasing adapter concentration beyond the recommended range may further improve ligation efficiency in some workflows.
Yes, contaminants such as salts, EDTA, ethanol, detergents, and extraction reagent carryover can impact enzymatic fragmentation. These may either inhibit or enhance enzyme activity, leading to inconsistent or shifted fragment sizes.
In some cases, fragmentation can be tuned by adjusting incubation time or temperature to compensate. However, if inhibitors significantly disrupt enzyme activity and no library is generated, we recommend performing an additional DNA cleanup before library preparation.
Handling and storage
Yes. The TrueAmp reagents have been tested through up to four freeze-thaw cycles with no observed impact on performance. These studies were conducted over multiple days to reflect typical laboratory use. As a best practice, repeated freeze–thaw cycles should still be minimized when possible.
The TrueAmp Library Preparation Kit has a shelf life of 12 months when stored as recommended. This is supported by accelerated aging studies conducted over 13 months, which showed no measurable impact on performance or stability.
Yes, we offer flexible customization options to support a wide range of specialized requirements or unique applications. For more information or to discuss your specific requirements, please contact our team.
Troubleshooting
The insert is the DNA of interest generated during fragmentation. The insert size refers to the length of this DNA region. The fragment (or sequencing fragment/library) includes the insert plus the sequencing adapters ligated to both ends. Therefore, fragment size = insert size + adapter sequences. Traditional adapter sequences are roughly 65-70nt long but dependent on the presence of UMIs or length of indexes. In paired-end sequencing, reads are generated from both ends of the insert. If the insert is longer than the combined read length, a portion in the middle remains unsequenced. If the insert is shorter, reads may overlap or extend into adapter sequences.
Insert size impacts library yield, sequencing performance, and downstream applications. Smaller fragments typically amplify more efficiently and can result in higher yields, while larger fragments may reduce cluster generation efficiency on some sequencing platforms. Insert size also influences data quality metrics such as coverage uniformity, duplication rates, and alignment efficiency.
In targeted enrichment workflows, especially multiplexed captures where the same mass input (e.g., 500 ng) represents a different number of molecules depending on fragment size. Smaller libraries contain more molecules than larger ones at the same mass, which can affect capture efficiency and sample balance.
Additionally, larger insert sizes in target enrichment workflows can increase the proportion of near-bait reads, as longer fragments extend beyond the probe-binding region. Insert size should also be considered relative to sequencing read length. For paired-end sequencing (e.g., 2×75 bp or 2×150 bp), inserts that are too short may result in overlapping reads or sequencing into adapter regions, reducing usable data. Selecting an appropriate and consistent insert size for your application helps ensure optimal sequencing performance and reliable downstream analysis.
Shorter-than-expected fragments typically result from degraded or pre-sheared input DNA, excessive fragmentation time or temperature, or buffer components (e.g., salts) that increase enzymatic activity. We recommend reducing fragmentation time or temperature and confirming input DNA quality to help achieve the desired insert size. If needed, performing a cleanup step prior to library preparation may improve consistency.
If your DNA has been sufficiently fragmented, the presence of multiple peaks, especially those appearing larger than the expected fragment size, is typically an artifact of library preparation rather than true size distribution:
1. PCR overamplification and heteroduplex formation. During excessive PCR cycling, DNA strands from different templates can anneal to each other, forming heteroduplexes. These structures migrate more slowly during capillary electrophoresis (e.g., TapeStation or Bioanalyzer), which can create the appearance of larger fragment sizes or additional peaks.
For additional information, please refer to the technical note on heteroduplexes.
2. Adapter structure effects (full-length Y adapters). Libraries constructed with full-length Y adapters (containing both P5 and P7 sequences) can exhibit altered migration behavior due to their partially single-stranded structure. This can cause fragments to appear larger than their true size during electrophoretic analysis.
These 2 effects can be validated by performing a short “reconditioning PCR”: Use a small number of cycles (e.g., 2–5 cycles) with the inclusion of a high concentration of primers. This step promotes the formation of fully matched homoduplexes, which migrate more accurately during electrophoresis. After this treatment, the apparent fragment size distribution should shift closer to the expected profile.
Low library yield can result from low input DNA, poor DNA quality, over- or under-fragmentation, or the presence of inhibitors that affect enzymatic steps. Additional factors such as improper size selection, incomplete ethanol removal, over-dried SPRI beads, or insufficient mixing during reactions may also reduce yield. Ensuring clean DNA input, optimizing fragmentation conditions, and following recommended SPRI handling and mixing practices should improve performance.
Adapter dimers or short fragments are typically observed with low input DNA, excess adapter concentration, or insufficient size selection. Additional causes include improper adapter handling (e.g., degradation or incorrect dilution) or pre-mixing adapters with ligation reagents. Optimizing adapter concentration, ensuring proper adapter handling, and performing appropriate SPRI cleanup (e.g., adjusting bead ratios or repeating cleanup) can help reduce these artifacts.
As a general starting point, the baseline fragmentation times and temperatures in the protocol are recommended to achieve insert sizes in the 200-250 bp range. Fragmentation behavior can vary depending on thermocycler calibration, DNA quality, and buffer composition. For best results, we recommend performing a small time-course optimization (DOE) using your specific samples. The full FAQ PDF includes a decision tree for optimizing the target insert size.
The number of PCR cycles depends on input DNA amount and desired library yield. Lower input samples typically require more cycles to generate sufficient material, while higher input samples require fewer cycles to minimize duplication and preserve library complexity. Optimization may be required to balance yield and sequencing performance.